Command Reference
R&S
®
ZNB/ZNBT
1061
User Manual 1173.9163.02 ─ 38
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM OFF
Disable the measurement in the created sweep segment.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:ADD
Inserts a new sweep segment using default channel settings ("Insert New Segment").
The added segment covers the frequency interval between the maximum frequency of
the existing sweep segments and the stop frequency of the entire sweep range.
Tip:
Use
[SENSe<Ch>:]SEGMent<Seg>:INSert
to create a segment with specific
channel settings.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. Segment numbers must be sequential.
If n segments exist already, the added segment must have the
segment number n+1.
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the
second segment will be between 1.5 MHz and the maximum fre-
quency of the analyzer.
Usage:
Event
Manual operation:
See
"Add / Insert / Delete / Delete All"
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
<ResBandwidth>
Defines the resolution bandwidth of the analyzer (Meas. Bandwidth) in sweep segment
no.
<Seg>
. Values between 1 Hz and 1 MHz can be set. Option R&S
ZNB/ZNBT-K17
enables receiver bandwidths up to 10 MHz (see
Chapter 4.7.4, "Receiver Bandwidth
Bandwidths can be set in 1
–
1.5
–
2
–
3
–
5
–
7 steps. The analyzer rounds up any
entered value between these steps and rounds down values exceeding the maximum
bandwidth.
At the same time, the command activates separate bandwidth setting in all sweep seg-
ments (
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
ON
).
Suffix:
<Ch>
.
Channel number
SCPI Command Reference