GUI Reference
R&S
®
ZNB/ZNBT
634
User Manual 1173.9163.02 ─ 38
Uncheck "Manual Configuration" if the device shall discover its host configuration via
DHCP. In this case only the device's "Hostname" can be specified.
In "Manual Configuration" mode also the unit's "IP Address", "Subnet Mask" and "Stan-
dard Gateway" can be set. If the unit is located in the same subnet as the analyzer,
simply use "Copy from local network" to copy the "Subnet Mask" and "Standard Gate-
way" from the analyzer's IP settings.
5.19.5.5
Switch Matrix RF Connections Dialog
The "Switch Matrix RF Connections" dialog allows to define the RF connections
between the VNA and the switch matrices and the RF connections to the (DUT) test
ports.
Access:
SYSTEM – SETUP > "External Ports" > "RF Connections..."
The following configuration rules apply:
●
Every VNA port must either be connected to a matrix port or directly assigned to a
DUT test port
●
Every matrix that was added to the connection configuration must be connected to
the VNA by at least one RF connection
●
Every VNA port that is not connected to a matrix must be assigned to a test port
●
The test ports must numbered consecutively, starting with 1
Configuration errors are displayed in balloon popups appearing at the lower left corner
of the dialog.
Setup Softtool