Command Reference
R&S
®
ZNB/ZNBT
696
User Manual 1173.9163.02 ─ 38
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Weight>
Weight relative to the "Cursor" tap
*RST:
0 dB
Default unit: dB
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:EYE:EMPHasis:STATe
<Boolean>
Activates/deactivates pre-emphasis in the calculation chain of the related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
*RST:
OFF
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:EYE:EQUalization:CTLE:DC
<DC Gain>
Specifies the DC gain of the CTLE (a two-pole filter with single zero) at the receiver
simulation of the related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
The equalizer can be enabled/disabled using
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<DC Gain>
*RST:
0 dB
Default unit: dB
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:EYE:EQUalization:CTLE:POLE<1|2>
<CTLE Poles>
Specifies the poles of the CTLE (a two-pole filter with single zero) used at the receiver
simulation of the related eye diagram.
SCPI Command Reference