Command Reference
R&S
®
ZNB/ZNBT
1111
User Manual 1173.9163.02 ─ 38
Query parameters:
<PowerCalSettingType>
WAVE | STARt | STOP | POINts | STYPe | ATTenuation |
CPOWer | CFRequency | TSTamp | LTSTamp | TVNA | MVNA |
MTESt | FSMode
WAVE
Calibrated port and calibration type (Gen=Source Power,
Rec=Measurement Receiver)
STARt
Start Frequency or Power
Stop Frequency or Power
POINts
Nr. of sweep points
STYPe
Sweep type or grid (LIN, LOG, SEGM)
ATTenuation
Source or receiver attenuation
CPOWer
Cal Power for frequency sweeps
CFRequency
Cal Frequency for power sweeps
TSTamp
Timestamp in UTC
LTSTamp
Timestamp in local time
TVNA
For calibrations involving switch matrices, this indicates the test
ports on the VNA itself. Returns a comma-separated list of port
pairs
TestPort,VnaPort
MVNA
For calibrations involving switch matrices, this indicates the
mapping between matrix ports and VNA ports for matrix
<Index>. Returns a comma-separated list of port pairs
MatrixVnaPort,VnaPort
MTESt
For calibrations involving switch matrices, this indicates the
mapping between matrix (physical) test ports and test ports for
matrix <Index>. Returns a comma-separated list of port pairs
MatrixTestPort,TestPort
FSMode
Returns the frequency sweep mode that was used during cali-
bration.
STEP:
Stepped mode (for all segments)
ANAL:
Swept mode (for at least one segment)
Because swept mode is limited to frequency sweeps this always
returns STEP.
SCPI Command Reference