[SENSe<Ch>:]CORRection:EDELay<PhyPt>:AUTO
<Activate>
Defines the offset parameter for the active test port such that the residual delay of the
active trace (defined as the negative derivative of the phase response) is minimized
across the entire sweep range ("Auto Length").
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer. This numeric suffix is ignored; the
active port is determined by the active trace.
Setting parameters:
<Activate>
ONCE
Applies the auto length function.
Example:
*RST; :CORR:EDEL:AUTO ONCE
Reset the instrument and apply the auto length function to the
default trace (Trc1 in channel 1).
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:DIELectric
<Permittivity>
Defines the permittivity for the offset correction at test port
<PhyPt>
.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<Permittivity>
Permittivity
Range:
1 to +1E+6
*RST:
1.00062
Example:
See
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:
See
"Permittivity / Velocity Factor"
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:DISTance
<MechLength>
Defines the offset parameter for test port
<PhyPt>
as a mechanical length.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer