GUI Reference
R&S
®
ZNB/ZNBT
361
User Manual 1173.9163.02 ─ 38
5.5.2.2
Port Settings Dialog
The "Port Settings" dialog configures the source and receive ports of the R&S
ZNB/
ZNBT for arbitrary frequencies, source and receiver levels. The source and receive fre-
quencies of the ports are always equal.
Access:
CHANNEL – CHANNEL CONFIG > "Port Config" > "Port Settings..."
Related information
Refer to the following sections:
●
Chapter 4.7.3, "Mixer and Frequency Conversion Measurements"
●
Chapter 5.19.4.2, "External Power Meters Dialog"
●
Chapter 5.19.4.4, "External Generators Dialog"
Arb Frequency Tab
Defines port-specific frequencies (arbitrary mode) and other port-specific source set-
tings.
Arbitrary mode requires software option R&S
and Frequency Conversion Measurements"
If any of the port frequencies are changed from their preset values, the frequency con-
version mode is activated on pressing "Apply" or "OK" in the dialog. An "Arb" label
appears in the channel line. "Reset Port Settings" plus "Apply" or "OK" deactivates the
arbitrary mode.
Channel base frequency
, all port frequencies are set to the channel base frequency
f
b
. For frequency sweeps, f
b
corresponds to the unmodified sweep range. For power,
time, and CW mode sweeps, it is equal to the CW frequency. f
b
is defined by the "Stim-
Chapter 5.8, "Stimulus Softtool"
Channel Config Softtool