Command Reference
R&S
®
ZNB/ZNBT
1041
User Manual 1173.9163.02 ─ 38
Suffix:
<Ch>
.
Channel number
Parameters:
<PortNumber>
Analyzer port number
Range:
1 to the number of analyzer ports
*RST:
2
Example:
See
[SENSe<Ch>:]FREQuency:IMODulation:LTONe
Manual operation:
See
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum:MORDer
<IMOrder>
Defines the maximum order of intermodulation products for the intermodulation spec-
trum measurement.
Suffix:
<Ch>
.
Channel number
Parameters:
<IMOrder>
Maximum order of IM products
Range:
3 | 5 | 7 | 9
*RST:
3
Example:
See
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:
Manual operation:
See
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:STATe]
<Boolean>
Enables or disables the measurement of the intermodulation spectrum. If a new chan-
nel is desired, the trace must be created separately; see second example below.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON | OFF - enables or disables intermodulation spectrum mea-
surement.
*RST:
OFF
SCPI Command Reference