GUI Reference
R&S
®
ZNB/ZNBT
544
User Manual 1173.9163.02 ─ 38
Target Format
Description
Formula
Imag
Imaginary part of z
Im(z) = y
SWR
(Voltage) Standing Wave Ratio
SWR = (1 + |z|) / (1 – |z|)
Default
Identical to trace format.
Note
: the Smith and Polar traces
use "Lin Mag" as the default for-
mat for target value.
-
Remote command:
CALCulate<Chn>:MARKer<Mk>:SEARch:FORMat
Target Search
Activates the search and sets the active marker to the defined target value. If the target
value occurs at several stimulus values, the marker is placed to the search result with
the smallest stimulus value. The other measurement points with the same target value
can be located using the "Search Right" function.
If the target is not found (e.g. because the active trace doesn't contain the target
value), then the active marker is not moved away from its original position.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
TARGet
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Search Left
Activates the search to the left of the active marker position and sets the active marker
to the defined target value. The target search range is between the "Start" value of the
search range and the active marker position.
If the target value occurs at several stimulus values, the marker is placed to the search
result with the largest stimulus value (i.e. the one which is closest to the active marker
position). The other measurement points with the same target value can be located
using "Search Left" repeatedly.
If the target is not found (e.g. because the active trace doesn't contain the target
value), then the active marker is not moved away from its original position.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
LTARget
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Search Right
Activates the search to the right of the active marker position and sets the active
marker to the defined target value. The target search range is between the active
marker position and the "Stop" value of the search range.
If the target value occurs at several stimulus values, the marker is placed to the search
result with the smallest stimulus value (i.e. the one which is closest to the active
marker position). The other measurement points with the same target value can be
located using "Search Right" repeatedly.
If the target is not found (e.g. because the active trace doesn't contain the target
value), then the active marker is not moved away from its original position.
Marker Softtool