The logic is identical to that of the single calibration case (see
5.4.1.9
SMARTerCal Wizard (Manual)
The "SmarterCal (Manual)" wizard guides you through the setup and execution of a
manual SMARTerCal.
Access:
CHANNEL – CAL > "Start Cal" > "SMARTerCal" – "Start... (Manual)"
The wizard proceeds through the following steps:
1.
"Ports and Type":
Select the ports to be calibrated, the calibration type to be used
and the power meter connection.
2.
"Connectors and Cal Kits":
Select the connector type and gender for all ports to
be calibrated. If necessary, load or change a calibration kit.
3.
"SMARTerCal":
a) Acquire error correction data for the required ports and port pairs and the
required standards
b) Acquire source power calibration data at the "Power Port"
c) Finally, decide whether or not to apply the resulting calibration
Background and related information
●
on page 172 for background information.
●
Chapter 5.4.1.10, "SMARTerCal Wizard (Manual, MultiCal) "
on page 314 for
the "MultiCal" variant of the wizard.
●
to perform an additional source flatness calibration (see
Chapter 4.5.7.3, "Combining SMARTerCal with Scalar Power Calibration"
●
If the active channel is already calibrated, the wizard will attempt to load the cali-
bration. If the calibration setup is unchanged and sweep data are available the
existing system error correction can be optimized without repeating the measure-
ment of all standards.
●
When you apply the acquired SMARTerCal, the active calibration is replaced and
discarded.
To persist any kind of calibration you can transfer it to the "Cal Pool" using the
●
The active system error correction data can be read (and modified) using the
remote control command
If external switch matrices are involved, use
Cal Softtool