Command Reference
R&S
®
ZNB/ZNBT
1010
User Manual 1173.9163.02 ─ 38
Timestamp in local time
TVNA
For calibrations involving switch matrices, this indicates the test
ports on the VNA itself. Returns a comma-separated list of port
pairs
TestPort,VnaPort
MVNA
For calibrations involving switch matrices, this indicates the
mapping between matrix ports and VNA ports for matrix
<Index>. Returns a comma-separated list of port pairs
MatrixVnaPort,VnaPort
MTESt
For calibrations involving switch matrices, this indicates the
mapping between matrix (physical) test ports and test ports for
matrix <Index>. Returns a comma-separated list of port pairs
MatrixTestPort,TestPort
CKIT
If created with FW version 2.30 or higher, for each port the name
of the used calibration kit is stored with the calibration.
The query can return one of the following:
the cal kit name
(string)
if available and unique
'Multiple'
if cal kit names are available but not unique, i.e. if <PhyPt> was
not specified and multiple cal kits were used (multi-port cals with
different connector types, merged cals)
'Unknown'
if cal kit names are not available (cals created with a FW version
< 2.30)
FSMode
Returns the frequency sweep mode that was used during cali-
bration.
STEP:
Stepped mode (for all segments)
ANAL:
Swept mode (for at least one segment)
<Index>
If one or more external switch matrices were used during cali-
bration, this refers to the index of a particular switch matrix (see
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:DEFine
Example:
SENSe:CORRection:DATA:PARameter?
Unrestricted query. Result looks like this:
100000,8500000000,201,-10,LIN
SENSe:CORRection:DATA:PARameter? STARt
...
SENSe:CORRection:DATA:PARameter? STYPe
Query settings one by one.
SENSe:CORRection:DATA:PARameter? RATTenuation
Query receiver attenuations. Result looks like this:
1,0.000000,2,0.000000
SCPI Command Reference