Parameters:
<Scale>
String parameter, one of the following signals:
'Channel Base; Source' - base channel power
'Port 1; Source' - source power at port 1
'Port 2; Source' - source power at port 2
...
'Gen 1; Source' - external generator power 1
...
*RST:
'Channel Base; Source'
Example:
*RST; SOUR:POW 5
Define a base channel power of +5 dBm. This power is also
used as an (initial) power for all test ports and external genera-
tors.
SOUR:POW1:OFFS -5, CPAD
Change the source power at test port no. 1 to 0 dB without
affecting the power at the remaining ports.
SWE:AXIS:POW 'Port 1; Source'
Select the source signal at port 1 as the reference signal for
power definitions.
SOUR:POW -5
Shift all powers by -5 GHz.
SWE:AXIS:POW 'Port 2; Source'
Select the source signal at port 2 as the reference signal for
power definitions.
SOUR:POW?
Query the power at test port 2. The response is 0.
Manual operation:
See
"Stimulus Axis – Frequency / Power"
[SENSe<Ch>:]SWEep:COUNt
<Sweep>
Defines the number of sweeps to be measured in single sweep mode
(
OFF
) and in channel no. <Ch>. Use
to define the sweep count for all channels.
Suffix:
<Ch>
.
Channel number
Parameters:
<Sweep>
Number of consecutive sweeps to be measured.
Range:
1 to 100000
*RST:
1
Example:
See
CALCulate<Chn>:DATA:NSWeep:FIRSt?
Manual operation:
See
SCPI Command Reference