GUI Reference
R&S
®
ZNB/ZNBT
366
User Manual 1173.9163.02 ─ 38
Attenuation
Sets the attenuation at the respective measurement receiver (b-wave). See
Remote command:
Common Controls
Modified "Port Settings" are applied on "Apply" or "OK". Use "Cancel" to discard possi-
ble changes.
Reset Port Settings
Resets the "Freq. Conversion" and "Power Conversion" settings of all ports. After
"Reset Port Settings", an "Apply" or "OK" terminates the arbitrary mode.
Remote command:
[SENSe<Ch>:]FREQuency:CONVersion
FUNDamental
Apply / OK
Activates or deactivates frequency conversion (arbitrary) mode, depending on the cur-
rent port configuration.
●
If individual port frequencies are selected, the arbitrary mode is activated.
●
If individual port powers but no individual frequencies are selected, the current
mode is maintained.
●
If all port frequencies are equal to the channel base frequency, and all port powers
are equal to the channel base power (e.g. after
, the arbitrary
mode is deactivated.
Remote command:
[SENSe<Ch>:]FREQuency:CONVersion
5.5.2.3
Define Parallel Measurement Dialog
With the R&S
ZNBT or a 4-port R&S
ZNB with internal second source you can mea-
sure mutliple DUTs (or multiple paths of a single DUT) in parallel. The "Define Parallel
Measurement" dialog allows to configure these DUTs and to declare the connections
Channel Config Softtool