Limit lines can be defined like the limit lines for time sweeps.
●
The bandfilter search functions are available for the transformed trace.
●
If marker coupling is active, then the markers in the time domain and in the fre-
quency domain are coupled with each other.
The analyzer places no restriction on the measured quantities to be transformed into
the time domain. Impedances and admittances are first converted back into the equiva-
lent S-parameter, transformed, and restored after the transformation.
See also
Chapter 4.7.2.1, "Chirp z-Transformation"
Remote command:
CALCulate<Chn>:TRANsform:TIME:STATe
Type
Selects a band pass or low pass time domain transform. See
To calculate a low pass transform the sweep points must be on a harmonic grid (other-
wise the analyzer will only be able to calculate an approximate result and generate a
warning). "Low Pass Settings..." opens a dialog that allows to establish or change a
harmonic grid (not available for memory traces).
See
Chapter 5.12.4.2, "Low Pass Settings Dialog"
Remote command:
CALCulate<Chn>:TRANsform:TIME[:TYPE]
CALCulate<Chn>:TRANsform:TIME:STIMulus
Impulse Response
Selects a window type which the R&S
ZNB/ZNBT uses to filter the trace in the fre-
quency domain. The drop-down list shows the impulse response of a constant trace
over a finite sweep range (i.e. a rectangular function) that was filtered using the differ-
ent available window types. The selected window is applied to the active trace.
See also
Chapter 4.7.2.3, "Windows in the Frequency Domain"
Note:
The frequency domain window is used to filter the trace prior to the time domain
transformation. An independent "Time Gate" can be used after the transformation in
order to eliminate unwanted responses (see
The analyzer always uses a "No Profiling (Rectangle)" window to calculate the time-
gated frequency domain trace, see
"Time-Gated Frequency Domain Trace"
Remote command:
CALCulate<Chn>:TRANsform:TIME:WINDow
Side Lobe Level
Defines the side lobe suppression for an "Arbitrary Sidelobes (Dolph-Chebychev)" win-
dow. The entered value is the ratio of the power of the central lobe to the power of the
first side lobe in dB.
Remote command:
CALCulate<Chn>:TRANsform:TIME:DCHebyshev