GUI Reference
R&S
®
ZNB/ZNBT
431
User Manual 1173.9163.02 ─ 38
Start Power / Stop Power
Defines the sweep range of for
sweeps.
Remote command:
SOURce<Ch>:POWer<PhyPt>:STARt
SOURce<Ch>:POWer<PhyPt>:STOP
Number of Points
Sets the total number of measurement points for
This value can also be set on the "Sweep Params" tab of the "Sweep"softtool (see
Remote command:
Start Time / Stop Time
Defines the duration of a
sweep.
Note:
The minimum accepted "Stop Time" may not be technically feasible. In this
case, an error message is displayed on sweep start.
Remote command:
Power
Determines the output power at the test ports for the sweep types "Lin Freq", "Log
Freq", "CW Mode", and "Time". Also determines the default output power for "Segmen-
ted" sweeps, where the output power can be set per segment.
The setting has no effect for "Power" sweeps, where the source power is varied over a
continuous range.
Remote command:
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
CW Frequency
Sets the fixed frequency for
,
, and
sweeps.
The "CW Frequency" is also used as the channel base frequency for frequency-con-
verting measurements; see
Chapter 4.7.3, "Mixer and Frequency Conversion Measure-
Remote command:
[SENSe<Ch>:]FREQuency:FIXed
[SENSe<Ch>:]FREQuency[:CW]
SOURce<Ch>:FREQuency<PhyPt>:FIXed
SOURce<Ch>:FREQuency<PhyPt>[:CW]
Zoom Stimulus
This function is only available, if the active trace is displayed in a diagram whose X
axis represents the stimulus.
Stimulus Softtool