GUI Reference
R&S
®
ZNB/ZNBT
425
User Manual 1173.9163.02 ─ 38
The remaining networks are defined by lumped elements. Their parameters are dis-
played below the graphical list. See
Chapter 4.6.2.4, "Circuit Models for 4-Port Net-
Tip:
●
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol. The lumped elements and S-parameter networks ("D1", "D2") are
numbered from top to bottom.
●
Conductance in Embedding Networks
switch to change from resistances
in "Capacitor in parallel with resistor" circuit blocks to conductances and vice versa.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:C<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:L<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:R<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>:
PARameters:G<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:C<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:L<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:R<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:G<Cmp>
D1, D2
The "D1" and "D2" buttons are enabled as long as a four-port network involving two-
port or four-port S-parameter data is selected. The networks are defined by their S-
parameters stored in a four-port or two-port Touchstone file (
*.s4p
,
*.s2p
). No addi-
tional parameters are required.
In case the port number conventions of the loaded Touchstone file differ from network
analyzer conventions (see description of
), it is possible to "Switch Gates".
The analyzer will interchange the port numbers (e.g. S
12
--> S
21
) when loading the file.
Note:
The loaded file is stored in the active recall set. Persisted recall sets contain the
full (de-)embedding data so that they can be transferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:BALanced:DEEMbedding<LogPt>
MMEMory:LOAD:VNETworks<Ch>:BALanced:EMBedding<LogPt>
Offset Embed Softtool