Example:
SENS1:AVER:COUN 15; :AVER ON
Set the average factor for channel 1 to 15 (the mnemonic
SENS1
can be omitted) and enable the sweep average.
AVER:COUN 5; CLE
Reduce the average factor and restart the average.
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]AVERage:COUNt
<AverageFactor>
Defines the number of consecutive sweeps to be combined for the sweep average
( "Factor" ).
Suffix:
<Ch>
.
Channel number
Parameters:
<AverageFactor>
Sweep average factor
Range:
1 to 1000
*RST:
10
Example:
See
Manual operation:
See
[SENSe<Ch>:]AVERage:MODE
<Mode>
Suffix:
<Ch>
.
Parameters:
<Mode>
AUTO | FLATten | REDuce | MOVing
AUTO
Automatic selection between REDuce and FLATten mode,
depending on the trace format.
FLATten
Cumulative moving averages of the (linear) magnitude and
phase values, provides the most effective noise suppression for
the "dB Mag" , "Phase" , "Unwr Phase" , and "Lin Mag" formats.
REDuce
Cumulative moving averages of the real and imaginary parts of
each measurement result, provides the most effective noise sup-
pression for the "Real" and "Imag" formats and for complex trace
formats.
MOVing
Simple moving averages of the real and imaginary parts of each
measurement result; similar to REDuce, but with finite history.
Manual operation:
See