Command Reference
R&S
®
ZNB/ZNBT
1121
User Manual 1173.9163.02 ─ 38
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number
<PhyPt>
.
*RST:
OFF
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient[:STATe]
<Boolean>
Enables or disables the use of two-port transmission coefficients.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
transmission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the two-port transmission coeffi-
cients.
*RST:
n/a
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:CALibration
<Boolean>
Selects the position of the additional two-port in the test setup.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
transmission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Parameters:
<Boolean>
ON
–
two port at power meter (during calibration)
OFF
–
two port at DUT (during measurement)
*RST:
ON (notice that the correction is switched off after a
reset)
SCPI Command Reference