GUI Reference
R&S
®
ZNB/ZNBT
336
User Manual 1173.9163.02 ─ 38
The accuracy of the source power calibration depends on the the power meter's mea-
surement accuracy, therefore it is advantageous to select a reference receiver cal
power at which the power meter provides a maximum accuracy. Otherwise your can
use equal port cal power and reference receiver cal power values.
Note: Risk of damage due to high power settings
If an external device (e.g. an amplifier) is connected between the calibrated test port
and the power meter, ensure that the reference receiver cal power does not exceed
the maximum input power of this device.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:PPOWer
SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect
5.4.3.3
Power Meter Transmission Coefficients Dialog
The "Power Meter Transmission Coefficients" dialog allows you to modify the scalar
power calibration data in order to account for an additional two-port device in the test
setup with known transmission coefficients. Configuration of the transmission coeffi-
cients and activation are independent from each other.
Access:
CHANNEL – CAL > "Pwr Cal Settings" > "Transm. Coefficients..."
Background information
Refer to
Chapter 4.5.6.4, "Extended Test Setups"
Controls in the Power Meter Transmission Coefficients Dialog
Test Setup
The button group on the right of the dialog allows to select a test setup with an addi-
tional two port in front of the DUT (during the measurement) or in front of the power
meter (during power calibration). "No Coefficients" disables the transmission coeffi-
cients but does not delete the entries in the
.
Cal Softtool