5.9.4 Intermod. Tab
Selects intermodulation products and derived results as measured quantities. The
intermodulation products occur at frequencies which correspond to sums and differen-
ces of the upper and lower tone frequencies and their integer multiples.
The intermodulation measurement requires options R&S
ZNB/ZNBT-K4 and R&S
ZNB/
ZNBT-K14. The results are available as soon as the measurement has been properly
configured on the "Intermod." tab of the Channel Config softtool (see
Background information
Refer to the following sections:
●
Chapter 4.7.3.3, "Intermodulation Measurements"
●
The labels for the intermodulation quantities in the trace list are identical to the param-
eters used for remote control; see
CALCulate<Ch>:PARameter:SDEFine
5.9.4.1
Controls on the Intermod. Tab
The following buttons in the "Intermod." tab open associated dialogs:
●
"More IM Products...": see
Chapter 5.9.4.2, "IM Products Dialog"
●
Chapter 5.9.4.3, "Intercept Points Dialog"
Meas Softtool