GUI Reference
R&S
®
ZNB/ZNBT
377
User Manual 1173.9163.02 ─ 38
Auto (column)
: Enables the automatic mode for the corresponding drive port, disa-
bling the manual configuration for the related a- and b-waves.
Reset
: restores the default settings ("Auto" for all drive ports)
Remote command:
All Segments – Reset
Restores the default settings for all segments and drive ports.
All Segments – Learn Sweep
During a learn sweep, the analyzer determines the appropriate static gain settings for
the measured a- and b-waves, i.e. for all a- and b-waves measured in the current
channel. The acquired settings can be overwritten manually.
At the start of the learn sweep, "Auto" mode is selected for each drive port ("Reset")
and a single shot measurement of the current channel is initiated.
During this measurement the VNA software observes the "Low Dist" (LD) vs. "Low
Noise" (LN) gain decisions of the AGC and derives the statically assigned gain for the
individual sweep segments / drive ports / receivers as follows:
●
if LD was selected for any of the related measurement points, then LD is assigned
●
otherwise LN is assigned
In other words, LN is assigned if and only if LN was selected for all related measure-
ment points.
Note:
●
Before running the "Learn Sweep", create the adequate port configuration and add
the required traces.
●
The "Learn Sweep" is not available for power sweep channels.
Remote command:
[SENSe<Ch>:]POWer:AGCMode:ACQuire
[SENSe<Ch>:]POWer:AGCMode:SAVE
Set All Items to ...
Convenience functions to apply the same manual AGC mode to all a and b wave
receivers.
If
is enabled, the setting only applies to the selected segment.
Remote command:
[SENSe<Ch>:]POWer:GAINcontrol:ALL
5.5.4 Intermod. Tab
Controls the intermodulation measurement including power calibration. The intermodu-
lation measurement requires options R&S
ZNB/ZNBT-K4 and R&S
ZNB/ZNBT-K14.
The results can be selected using the "Intermod." tab of the "Meas" softtool, see
Intermodulation Presetting Wizard
guides through the measurement configuration.
Channel Config Softtool