GUI Reference
R&S
®
ZNB/ZNBT
284
User Manual 1173.9163.02 ─ 38
The wizard proceeds through the following steps:
1.
"Ports and Type":
Define the sets of ports to be calibrated and the calibration
types to be used for the individual sets
2.
"Connectors and Cal Kits":
Select the connector type, gender and cal kit for all
ports to be calibrated. If necessary, import a calibration kit.
3.
"Calibration":
Acquire measurement data for the required ports or port pairs and
the required standards. Finally, decide whether or not to apply the resulting system
error correction.
Background and related information
●
The "Multiple Cal in Calibration Wizard" feature is disabled by default and has to be
activated in the
of the System Config dialog.
●
Chapter 5.4.1.4, "Calibration Presetting Wizard"
information.
Step 1: Ports and Type
The first page of the wizard presents the list of already configured calibrations and
allows to
●
"Add"new calibrations
●
"Delete" or "Modify" existing calibrations
Calibrate all Channels
Check this box to apply the acquired correction data to all channels in the active recall
set. Leave it unchecked (preset setting) to apply them only to the active channel.
Note that this option is available only if the active recall set contains multiple channels.
Remote command:
[SENSe:]CORRection:COLLect:CHANnels:ALL
Cal Softtool