Parallel Measurement with Frequency Offset
Enables/disables a frequency offset in parallel measurements.
For segmented sweeps the parallel measurements are performed segment per seg-
ment. If the
is specified as a "Bandwidth Factor" and the segments
use different measurement bandwidths, this may result in different frequency offsets for
different segments.
Note:
If both "Parallel Measurement with Frequency Offset" and
active, it may happen that automatic switches between small and large IF gain during
the sweep may lead to visible steps in the measurement results. This can be avoided
by selecting a fixed AGC mode ("Low Dist" or "Low Noise").
Remote command:
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:STATe
Minimum Offset
The "Minimum Offset" can be specified either as an absolute value ("Direct Value") or
as a multiple of the measurement bandwidth ("Bandwidth Factor").
Remote command:
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:MODE
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:DVALue
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:BWFactor
State indication warning
When attempting to activate the "Parallel Measurement with Frequency Offset" (i.e. on
closing the "Define Parallel Measurement" dialog with the "OK" button) the VNA Firm-
ware checks whether the requested settings are technically feasible. If this is not the
case, the firmware disables parallel measurement (i.e. the measurements are actually
performed one after the other) and generates a warning message indicating the current
configuration problem.
In particular, activation is rejected if
●
the start frequency is < 31 MHz
●
the measurement bandwidth is > 100 kHz
●
the resulting frequency offset (i.e. the minimum frequency offset, rounded to a mul-
tiple of the current frequency step size) would be too high
●
the port groups are configured with different frequency conversion settings
The generated error code can be retrieved via remote control command
SOURce<Ch>:GROup:SIMultaneous:FOFFset:CONDition?
. The configuration
problems listed above are indicated as error codes -8 "invalid offset or frequency spac-
ing" and -6 "no simultaneous mode possible".
Remote command:
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:CONDition