SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:PMReadings
<NoReadings>
Selects the number of power meter readings for source power calibration method
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:METHod RRAFter
.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
Parameters:
<NoReadings>
Number of power meter readings. If the value exceeds the maxi-
mum number of calibration sweeps
(
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
AVERage[:COUNt]
) the latter is adjusted.
Range:
1 to 100
*RST:
1
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Gets the number of available power calibrations (both source power and measurement
receiver).
Use
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
to retrieve the
related settings.
Suffix:
<Ch>
.
Number of the calibrated channel.
<PhyPt>
This suffix is ignored.
<Wv>
This suffix is ignored.
Usage:
Query only
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe]
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for an external
generator number
<Gen>
. The command is disabled unless a source power calibration
for the external generator has been performed (
). To enable or disable a source power calibration for an
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
.
Suffix:
<Ch>
.
Calibrated channel number
SCPI Command Reference