Command Reference
R&S
®
ZNB/ZNBT
784
User Manual 1173.9163.02 ─ 38
CALCulate<Chn>:STATistics:DOMain:USER
<EvalRange>
Selects one out of 10 evaluation ranges to be configured with the
CALCulate<Chn>:STATistics:DOMain:
, and
CALCulate<Chn>:STATistics:DOMain:USER:STOP
mands.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<EvalRange>
Number of the evaluation range.
Range:
1 to 10. In addition, 0 denotes the (non-configura-
ble) "Full Span" evaluation range.
*RST:
0
Example:
*RST; :CALC:STAT:DOM:USER?
Query the default evaluation range. The response is zero, i.e.
the evaluation range is equal to the complete sweep range
CALC:STAT:DOM:USER 1
CALC:STAT:DOM:USER:STARt 1GHZ; STOP 2GHZ; SHOW
ON
Select evaluation range no. 1 and define the evaluation range
between 1 GHz and 2 GHz. Display the range limit lines.
Manual operation:
See
CALCulate<Chn>:STATistics:DOMain:USER:SHOW
<Boolean>
Displays or hides range limit lines for the evaluation range selected via
CALCulate<Chn>:STATistics:DOMain:USER
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - range limit lines on or off.
*RST:
OFF
Example:
See
CALCulate<Chn>:STATistics:DOMain:USER
Manual operation:
See
CALCulate<Chn>:STATistics:DOMain:USER:STARt
<Start>
CALCulate<Chn>:STATistics:DOMain:USER:STOP
<Stop>
These commands define the start and stop values of the evaluation range selected via
CALCulate<Chn>:STATistics:DOMain:USER
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI Command Reference