Command Reference
R&S
®
ZNB/ZNBT
786
User Manual 1173.9163.02 ─ 38
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Stop>
Stop value of the evaluation range.
Range:
+0.01 dBm to +100.00 dBm
*RST:
1 dBm
Default unit: dB
Example:
See
CALCulate<Chn>:STATistics:NLINear:COMP:
Manual operation:
See
CALCulate<Chn>:STATistics:NLINear:COMP:RESult?
Returns the x-dB compression point of an S-parameter or ratio measured in a power
sweep. The compression value x is set via
.
The response contains two numeric values:
●
<Cmp In>
–
stimulus level at the compression point in dBm.
●
<Cmp Out>
–
sum of <Cmp In> plus the magnitude of the measured response
value at the compression point in dBm.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
*RST; SWE:TYPE POW
Select a power sweep with default CW frequency and sweep
range.
CALC:STAT:NLIN:COMP:LEV 2
Define a compression value of 2 dB.
CALC:STAT:NLIN:COMP:RES?
Query the compression point results <Cmp In>, <Cmp Out>. An
execution error message (error no. -200) is returned if no com-
pression point is found.
CALC:STAT:NLIN:COMP ON
Display the compression point result in the diagram area.
Usage:
Query only
Manual operation:
See
CALCulate<Chn>:STATistics:NLINear:COMP[:STATe]
<Boolean>
Displays or hides the compression point result in the diagram area of trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI Command Reference