GUI Reference
R&S
®
ZNB/ZNBT
309
User Manual 1173.9163.02 ─ 38
Remote command:
The port parameters in many calibration commands define the calibrated port(s).
Type
Selects the calibration type to be performed. For background information see
ter 4.5.7.2, "Calibration Types"
SMARTerCal is based on a full n-port calibration (n > 1): "Adapter Removal" is a 2-port
calibration, all other calibration types can be used for any combination of 2 or more
ports.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID
Power Port
Selects the port for the scalar source power calibration.
Calibrate all Channels
Check this box to apply the acquired correction data to all channels in the active recall
set. Leave it unchecked (preset setting) to apply them only to the active channel.
Note that this option is available only if the active recall set contains multiple channels.
Remote command:
[SENSe:]CORRection:COLLect:CHANnels:ALL
Select All/Deselect All
Selects/deselects all ports.
These buttons are only shown if more than 4 ports are available.
Next
"Step 2: Connectors and Cal Kits"
on page 309. Inactive (and a warning
is displayed) as long as and.
Step 2: Connectors and Cal Kits
Selects the connector type and gender for all ports and allows you to import a calibra-
tion kit.
Background information
Refer to
Chapter 4.5.2, "Calibration Standards and Calibration Kits"
Cal Softtool