Command Reference
R&S
®
ZNB/ZNBT
1133
User Manual 1173.9163.02 ─ 38
Parameters:
<Boolean>
ON | OFF - turns the internal source power at the specified test
port no.
<PhyPt>
on or off.
*RST:
ON
Example:
*RST; SOUR:POW -6
Set the internal source power for channel 1 and all test ports to
-6 dBm.
SOUR:POW2:STAT?
Query whether the source power at test port 2 is on. The ana-
lyzer returns a
1
.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:SWEepend:MODE
<SweepEndModes>
Selects the power mode at sweep end.
This is a global setting.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the setting is valid for all
channels.
<PhyPt>
Physical port number. This suffix is ignored; the setting is valid
for all ports.
Parameters:
<SweepEndModes> AUTO | REDuce | KEEP
AUTO: at sweep end restore the power at sweep start
REDuce: at sweep end reduce the output power as if the chan-
nel base power was set to its minimum possible value
KEEP: at sweep end keep the power at its current level
Example:
SOURce:POWer:SWEepend:MODE REDuce
Reduce power at sweep end.
SOURce:POWer:SWEepend:SDELay 0.01
Use a settling delay of 10 ms
SOURce:POWer:SWEepend:MODE AUTO
At sweep end, restore the power at sweep start
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:SWEepend:SDELay
<SettlingDelay>
If sweep end mode REDuce or KEEP is active (see
) this defines the settling delay at subsequent sweep starts, i.e. the
time between power reset and sweep start when a new sweep is requested.
This is a global setting.
SCPI Command Reference