Command Reference
R&S
®
ZNB/ZNBT
1063
User Manual 1173.9163.02 ─ 38
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<Selectivity>
NORMal | MEDium | HIGH
NORMal - IF filter with normal selectivity and shortest settling
time.
MEDium - IF filter with steeper edges and longer settling time.
HIGH - IF filter with highest selectivity but longest settling time.
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol
<Boolean>
Selects common or independent "Selectivity" settings for the individual sweep seg-
ments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.
Parameters:
<Boolean>
ON - use independent selectivity settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:
SELect
OFF - reset the selectivity in all sweep segments to the selectiv-
ity for unsegmented sweeps, defined via
previous values.
The parameter is automatically switched to ON when a selectiv-
ity is entered using
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
*RST:
OFF
Example:
*RST; :SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus NORMal selectivity.
SEGM:BWID:SEL HIGH
Change the selectivity to HIGH.
BWID:SEL?
Query the (default) selectivity for unsegmented sweeps. The
response is NORM.
SEGM:BWID:SEL:CONT OFF
Couple the selectivities in all segments and reset the selectivity
in segment no. 1 to the unsegmented value NORMal.
SCPI Command Reference