Command Reference
R&S
®
ZNB/ZNBT
960
User Manual 1173.9163.02 ─ 38
7.3.15.2
[SENSe:]BANDwidth...
The
[SENSe:]BANDwidth...
commands set the bandwidth of the IF measurement
filter (resolution bandwidth).The forms
BANDwidth
and
BWIDth
are equivalent.
[SENSe<Ch>:]BANDwidth[:RESolution]
............................................................................960
[SENSe<Ch>:]BWIDth[:RESolution]
.................................................................................960
[SENSe<Ch>:]BANDwidth[:RESolution]:SELect
................................................................ 960
[SENSe<Ch>:]BWIDth[:RESolution]:SELect
......................................................................960
[SENSe<Ch>:]BANDwidth[:RESolution]
<ResBandw>
[SENSe<Ch>:]BWIDth[:RESolution]
<ResBandw>
Defines the resolution bandwidth of the analyzer (measurement bandwidth). Values
between 1 Hz and 1 MHz can be set. Option R&S
ZNB/ZNBT-K17 enables receiver
bandwidths up to 10 MHz (see
Chapter 4.7.4, "Receiver Bandwidth 10 MHz"
Bandwidths can be set in 1
–
1.5
–
2
–
3
–
5
–
7 steps. The analyzer rounds up any
entered value between these steps and rounds down values exceeding the maximum
bandwidth.
Suffix:
<Ch>
.
Channel number
Parameters:
<ResBandw>
Resolution bandwidth
Range:
See above
Increment: 1-1.5-2-3-5-7 steps
*RST:
10 kHz
Default unit: Hz
Example:
BAND 1.1
Set a resolution bandwidth of approx. 1.1 Hz for channel 1.
BAND?
The analyzer returns the rounded bandwidth of 1.5 Hz.
Manual operation:
See
[SENSe<Ch>:]BANDwidth[:RESolution]:SELect
<Selectivity>
[SENSe<Ch>:]BWIDth[:RESolution]:SELect
<Selectivity>
Defines the selectivity of the IF filter for an unsegmented sweep. The value is also
used for all segments of a segmented sweep, provided that separate selectivity setting
is disabled (
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:
CONTrol OFF
).
Suffix:
<Ch>
.
Channel number
SCPI Command Reference