Command Reference
R&S
®
ZNB/ZNBT
969
User Manual 1173.9163.02 ─ 38
Example:
MMEM:LOAD:CKIT 'C:\Users\Public
\Documents\Rohde-Schwarz
\Vna\Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file
New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:N50:SEL 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1.calkit
reads
New
Kit 1
.
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:<StandardType>
<ConnType>, <CalKitName>,
<StandardLabel>, <MinFreq>, <MaxFreq>[, <DelayParam>, <Loss>, <Z0>[,
<C0>, <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>[, OPEN | SHORt | MATCh |
<Resistance>[, <Port1>[, <Port2>]]]]]
[SENSe:]CORRection:CKIT:<StandardType>?
<ConnType>, <CalKitName>[,
<Port1>[, <Port2>]]
Defines the parameters of a (possibly non-ideal) 1 port or 2-port calibration standard
<StandardType>
within a particular cal kit. Depending on the standard type, only a
subset of the parameters may be used; see
Suffix:
<StandardType>
.
<string>
Standard type
For one-port standards the first character denotes the gender,
for transmission standards the first two characters denote the
genders on both ends, e.g.
MOPen
for a male Open standard or
FFTHrough
for a Through standard with female connectors.
For a complete list of standards refer to
Parameters:
<StandardLabel>
Additional string label for the standard, typically the standard's
serial number
CalStandardProps
Parameters <MinFreq>, ..., OPEN | SHORt | MATCh | <Resist-
ance> define the properties of the calibration standard. See
.
Note:
Sliding Match and Attenuation standards have only 2
parameters (<MinFreq>,<MaxFreq>), Through and Line stand-
ards only have 5 parameters (<MinFreq>,...,<Z0>).
SCPI Command Reference