Command Reference
R&S
®
ZNB/ZNBT
970
User Manual 1173.9163.02 ─ 38
<Port1>, <Port2>
Optional port restriction.
For a one-port standard the validity of the standard characteriza-
tion can be restricted to a single port, for a two-port standard it
can be restricted to a port pair (specified using ascending port
numbers).
Note:
with a port restriction the defined standard becomes sex-
less. Hence, for each port (pair) there can be only one standard
of a given type, i.e.
:SENSe:CORRection:CKIT:MOP
<ParameterList>,1
and
:SENSe:CORRection:CKIT:FOP
<ParameterList>,1
will overwrite each other.
Parameters for setting and query:
<ConnType>,
<CalkitName>
String parameters uniquely identifying the cal kit to which the
standard belongs.
Note:
If the specified cal kit does not exist, it is created with the
specified calibration standard.
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Define the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
CORR:CKIT:FOP? 'N 50 Ohm'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in the active cal kit.
CORR:CKIT:FOP? 'N 50 Ohm','New Kit 1'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
Manual operation:
See
"One Port Standards / Two Port Standards"
Table 7-9: Set command parameters
One port standards
Two port standards
Open
(Offset)
Short
Match
Reflect
Sliding
match
Symmetric
Network
Through,
Line
Attenuation
<Con-
nType>
mandatory
<CalKit-
Name>
mandatory (and must not be an empty string)
<Standar-
dLabel>
mandatory (but may be an empty string)
<MinFreq>,
<MaxFreq>
mandatory
<Delay-
Param>,
<Loss>,
<Z0>
mandatory
mandatory
mandatory
mandatory
not used
mandatory
not used
<C0>,
<C1>,
<C2>, <C3>
mandatory
mandatory
*
mandatory
*
mandatory
**
mandatory
**
not used
SCPI Command Reference