GUI Reference
R&S
®
ZNB/ZNBT
324
User Manual 1173.9163.02 ─ 38
Related information
Refer to the following sections:
●
Chapter 4.5.2.4, "Cal Kit Files"
●
Chapter 5.4.2.2, "Calibration Kits Dialog"
●
Chapter 4.5.2.1, "Calibration Standard Types"
Access:
> "Standards..."
One port and two port standards are listed in two separate tables. Most of the buttons
on the right side are available only if the "Kit Standards" dialog was opened for a user-
defined calibration kit.
One Port Standards / Two Port Standards
The standard tables contain the following information:
●
"Type" and "Gender" describe the calibration standard type; for an overview see
Chapter 4.5.2.1, "Calibration Standard Types"
●
"Label" is a user defined name of the standard. The label can help you identify a
standard or distinguish different standards with similar parameters.
●
"Min Freq" and "Max Freq" define the rated frequency range of the standard. Dur-
ing calibration, the analyzer checks whether the sweep range is within the validity
range of all measured standards and possibly generates a warning.
●
".s1p File" and ".s2p File" define whether the characteristics of the standard are
described by a Touchstone file rather than by a circuit model from which the
R&S
ZNB/ZNBT can calculate the S-parameters. See
"View / Modify Cal Kit Standards Dialog"
Cal Softtool