background image

HI758

  

Marine Calcium

INSTRUCTION MANUAL

Functional Description

Errors and Warnings

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Under Range:

 A blinking “200” indicates that the reading is under 

range.

Over Range:

 A blinking value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Inverted Cuvette:

 The sample and the zero cuvette are inverted.

Low Battery:

 The battery is low and must be replaced soon. 

Dead Battery:

 The battery is dead and must be replaced. Once this 

indication is displayed, normal operation of the instrument will be 
interrupted. Change the battery and restart the meter.

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. 

Please check the preparation of the zero cuvette.

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument.  
For more information about Hanna Instruments and our products, visit 
www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us 
at [email protected]
Find your local Hanna Instruments Office on www.hannainst.com.

Range

200 to 600 ppm

Resolution

1 ppm

Accuracy

±6 % of reading @ 25 ˚C / 77 ˚F

Light Source

light emitting diode @ 610 nm

Light Detector

silicon photocell

Method

adaptation of the zincon method

Environment

0 to 50 °C (32 to 122 °F); RH max 95% non-condensing

Battery Type

1.5V AAA (1 pc.)

Auto-off

after 10 minutes of non-use

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

64 g (2.3 oz)

Specifications

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office. 
Each 

HI758

 meter is supplied complete with:

•  Sample Cuvettes and Caps (2 pcs.)
•  Reagents for 25 tests
•  1 mL syringe with tip 
•  minipipette with tips
•  plastic pipette
•  1.5V AAA battery
•  Instruction manual and quick reference guide

Preliminary Examination

Reviews: