Getting Started
Transmission Measurements
Operating manual 1303.6538.62 - 05
49
6. Press
Device Setup
again
and select a file format or
Clipboard.
7. Close the dialog and press
Screen
again to copy the diagram to
a file or an external application.
8. Open the
Nwa-File
menu and select
Save NWA As..
.
9. In the
Save As
dialog opened, select a file location, format and name
and activate
Save.
The active setup is stored to a file and can be reused in a later session.
Proceed as described in section
Starting the Analyzer, Shutdown
to shut down your
analyzer.
2.2 Transmission Measurements
A transmission measurement involves the same steps as a reflection measurement.
Note the following differences:
J
The test setup for transmission measurements involves two DUT and analyzer
ports. You can connect the input of your DUT to port 1 of the analyzer, the output to
port 2. After a preset, the analyzer will measure the forward transmission S-
parameter S
21
.
J
The analyzer provides special calibration types for transmission measurements.
Use the calibration wizard and select an appropriate type. A TOSM calibration will
correct the system errors for all transmission and reflection S-parameters.