![Rohde & Schwarz ZVL series Operating Manual Download Page 536](http://html.mh-extra.com/html/rohde-and-schwarz/zvl-series/zvl-series_operating-manual_1477904536.webp)
R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
521
set: the value must be 0.
Range [def. unit]
0 s to 2.5E+003 s. [s]
AUTO
activates automatic sweep time setting in the
segment, which is equivalent to a meas. delay of 0 s.
<Unused>
Ignored parameter, should be set to the default value 0.
<Meas.
Bandwidth>
Resolution bandwidth in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
.
Range [def. unit]
1.0E-6 Hz to 5 MHz [Hz].
<LO>
Position of the local oscillator frequency LO relative to the RF frequency
(Spur
Avoid)
. In remote control this parameter must be set when a sweep segment is
created.
Range [def. unit]
POSitive: LO > RF
SCPI,
Command
Types
Device-specific, with query (used to retrieve the channel settings for a particular
sweep segment).
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings.
SEGM:DEF?
Query the channel settings for the new segment.
[SENSe<Ch>:]SEGMent<Seg>:DELete
Deletes the specified (single) sweep segment
(Del. Selected Segment)
.
[SENSe<Ch>:]SEGMent<Seg>:DELete:ALL
deletes all segments in the channel.
<Ch>
Channel number.
<Seg>
Sweep segment number. If unspecified, the value is set to 1.
*RST value
–
SCPI, Command
Types
Device-specific, no query.
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and
thus NORMal selectivity.
SEGM:DEL
Delete the created segment
.
[SENSe<Ch>:]SEGMent:DELete:ALL
Deletes all sweep segments in the channel
(Del. All Segments)
.
[SENSe<Ch>:]SEGMent<Seg>:DELete
deletes a single segment.
<Ch>
Channel number.
*RST value
–