R&S ZVL
GUI Reference
SETUP Key
Operating Manual 1303.6580.32-06
295
Remote control:
SYSTem:LXI:MDEScription
LAN Reset
Initiates the network configuration reset mechanism (Lan Configure Initialize, LCI) for the instrument.
Default state of the network settings
According to the LXI standard, an LCI must place the following parameters to a default state.
Parameter Value
TCP/IP Mode
DHCP + Auto IP Address
Dynamic DNS
Enabled
ICMP Ping
Enabled
Password for LAN configuration
LxiWebIfc
The LCI for the network analyzer also resets the following parameters:
Parameter Value
Description
Rohde & Schwarz Vector Network Analyzer / ZVL / <serial no.>
Negotiation
Auto Detect
VXI-11 Discovery
Enabled
The LAN settings are configured using the instrument's LXI Browser Interface. With firmware versions
V3.00 and higher, a user account with administrator rights is necessary for the LXI functionality (see
Operation with and without Administrator Rights
in chapter 1).
Remote control:
SYSTem:LXI:LANReset
LXI Browser Interface
LAN eXtensions for Instrumentation (LXI) is an instrumentation platform for measuring instruments and
test systems that is based on standard Ethernet technology. LXI is intended to be the LAN-based
successor to GPIB, combining the advantages of Ethernet with the simplicity and familiarity of GPIB.
LXI classes and LXI functionality
LXI-compliant instruments are divided into three classes, A, B and C, with the functionality of the classes
hierarchically based one upon the other: