R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
522
SCPI, Command
Types
Device-specific, no query.
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings
and thus NORMal selectivity.
SEGM:ALL
Delete the created segment and all segments in the channel created before.
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:CENTer?
Returns the center frequency of sweep segment no. <Seg>.
<Ch>
Channel number.
<Seg>
Sweep segment number.
Response
Center frequency of the sweep.
Range [def. unit]
Depending on the instrument model [Hz].
*RST value
–
SCPI, Command
Types
Device specific, query only.
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ
Create a sweep segment with a sweep range between 1.0 MHz and 1.5
MHz.
SEGM:FREQ:CENT?
Query the center frequency of the created segment. The response is
1.2500006.
Note: The frequency range of the sweep segment can be changed via
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP
.
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN?
Returns the width of the frequency range of sweep segment no. <Seg>.
<Ch>
Channel number.
<Seg>
Sweep segment number.
Response
Frequency span of the sweep.
Range [def. unit]
Depending on the instrument model [Hz].
*RST value
–
SCPI, Command
Types
Device specific, query only.
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ
Create a sweep segment with a sweep range between 1.0 MHz and 1.5