R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
491
SCPI,
Command
Types
Device-specific, command or query
Example:
MMEM:LOAD:CKIT
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit'
Load the previously created cal kit file New_kit.calkit from
the default cal kit directory.
C
ORR:CKIT:N50:SEL 'New_kit'
Assign the imported kit to the N 50
V
connector type (assuming that the cal kit
name stored in
New_kit.calkit
reads New_kit).
Cal kit naming conventions
Calibration kit names must be entered as string parameters. The string contains the cal kit name used in
the
Calibration Kits
dialog; a
V
in the name must be replaced by 'Ohm', e.g.:
'NewKit1' denotes the user-defined calibration kit
NewKit1
.
'N 50 Ohm Ideal Kit' denotes the
N 50
V
Ideal Kit
.
'ZV-Z21 typical' denotes the cal kit
ZV-Z21 typical
.
[SENSe<Ch>:]CORRection:CKIT:SELect '<conn_name>', '<ckit_name>'
Selects the calibration kit to be used for a specified connector type <conn_type>.
The command is suitable for connector types with arbitrary, user-defined names. For standard
connector types you can use the command
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect
.
<Ch>
Channel number. This suffix is ignored because calibration kits are channel-
independent.
'<conn_name>'
Connector type, e.g. a user-defined connector type (string variable)
'<ckit_name>'
String parameters containing the name of a calibration kit available on the
analyzer.
*RST value
– (A *RST does not change the assignment between connector types and
calibration kits.)
SCPI,
Command
Types
Device-specific, command or query (the query requires the first string parameter
only)
Example:
MMEM:LOAD:CKIT
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit'
Load the previously created cal kit file New_kit.calkit from
the default cal kit directory.
C
ORR:CKIT:SEL 'N 50 Ohm', 'New_kit'
Assign the imported kit to the N 50
V
connector type (assuming that the cal kit
name stored in
New_kit.calkit
reads New_kit).