<Ph_pt>
Physical port number
<no>
Number of resistance in circuit model. If unspecified the numeric suffix is set to
1.
Range [def. unit]
1 to 3 [–] The number of resistances depends on the selected circuit model.
Parameters
Possible circuit models (character data); see circuit models.
*RST value
– (see command type description below)
<numeric_value>
Resistance R<no> for the specified circuit model.
Range [def. unit]
–10 M
V
to 10 M
V
[
V
]. The increment is 1 m
V
(10
–3
V
).
*RST value
0
V
for all resistances connected in series with an inductance
10 M
V
for all resistances connected in parallel with a capacitance
SCPI, Command
Types
Device-specific, command or query. In the query form, the <numeric_value>
parameter must be omitted. The command returns the resistance value for the
specified circuit model.
Example:
*RST; CALC:TRAN:VNET:SEND:EMB:PAR:R1? CSL; R2? CSL
Query the default resistances for the
Serial C, shunt L
circuit model. The
response is 10000000; 0.
CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3
Increase the resistance R2 to 2.2 k
V
.
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>[:STATe]
<Boolean>
Enables or disables the embedding function. It is allowed to change the circuit model and its parameters
while embedding is enabled.
<Ch>
Channel number.
<Ph_pt>
Physical port number
<Boolean>
ON – Embedding active
OFF – Embedding inactive
*RST value
OFF [–]
SCPI, Command
Types
Device-specific, command or query
Example:
CALC:TRAN:VNET:SEND:EMB:TND CSL
Select the
Serial C, shunt L
circuit model for embedding.
CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3; CALC:TRAN:VNET:SEND:EMB ON
Increase the resistance R2 for the
Serial C, shunt L
circuit model to 2.2 k
V
and enable embedding.
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:TNDefinitio
n FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL
Selects the circuit model for sembedding.