R&S ZVL
Preparing for Use
Putting the Instrument into Operation
Quick Start Guide 1303.6538.62 - 06
21
1.3 Putting the Instrument into Operation
This section describes the basic steps to be taken when setting up the analyzer for the
first time.
General safety instructions
Before turning on the instrument, please make sure that the following conditions are
fulfilled:
C
Instrument covers are in place and all fasteners are tightened.
C
Fan openings are free from obstructions.
C
Signal levels at the input connectors are all within the specified maxima.
C
Signal outputs are correctly connected and not overloaded.
C
The instrument is dry and shows no condensation.
Non-observance may cause damage to the instrument!
1.3.1 Unpacking the Instrument and Checking the Shipment
The instrument is shipped together with its mandatory accessories in a cardboard box.
In order to unpack its contents proceed as follows:
1. Open the cardboard box.
2. Remove the accessories packed into the box and take the instrument out of the
packaging.
3. Check the shipment against the list of accessories to ensure that all items are
included.
4. Remove the two protective caps from the front and rear and carefully inspect the
analyzer to make sure that it was not damaged during shipment.
Should the instrument be damaged, immediately notify the forwarder who shipped the
instrument to you and keep the container and packing material.
Equipment returned or sent in for repair must be packed in the original container or
packing with electrostatic protection. It is recommended to keep at least the two
protective caps for front and rear side in order to prevent damage to the controls and
connectors.