![Rohde & Schwarz ZVL series Operating Manual Download Page 475](http://html.mh-extra.com/html/rohde-and-schwarz/zvl-series/zvl-series_operating-manual_1477904475.webp)
R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
460
C:\R_S\Instr\user\Nwa\CableTypes
for user-defined cables) are not affected; see example.
'<source_directory>'
String parameter to specify the source directory.
*RST value
–
SCPI, Command Types
Device-specific, no query.
Example:
See MMEMory:STORe:CABLe
MMEMory:LOAD:CKIT '<file_name>'
Loads cal kit data from a specified NWA cal kit file.
'<file_name>'
String parameter to specify the name and directory of the cal kit file to be loaded. If
no path is specified the analyzer searches the current directory, to be queried with
MMEMory:CDIRectory?
.
Note: The loaded file must be a NWA-specific cal kit file with the extension *.calkit.
ZVR cal kit files can be imported using the
[SENSe<Ch>:]CORRection:CKIT:INSTall
command. Agilent cal kit
files can be imported manually and converted into *.calkit files.
*RST value
–
SCPI,
Command
Types
Device-specific, no query.
Example:
MMEM:LOAD:CKIT
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit'
Load the previously created cal kit file New_kit.calkit from
the default cal kit directory.
... :MMEM:STOR:CKIT 'New_kit',
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit'
Store the data for the user-defined cal kit Newkit and
overwrite the cal kit file New_kit.calkit.
MMEMory:LOAD:CKIT:SDATa '<conn_name>', '<ckit_name>', MMTHrough |
MFTHrough | FFTHrough | MMLine | MFLine | FFLine | MMATten | MFATten |
FFATten | MMSNetwork | MFSNetwork | FFSNetwork | MOPen | FOPen | MSHort |
FSHort | MOSHort | FOSHort | MREFlect | FREFlect | MMTCh | FMTCh | MSMatch |
FSMatch, '<stdlabel_name>', '<file_name>' [,<port1_no>][,<port2_no>]
Loads cal kit data for a specific connector type, cal kit, and calibration standard from a specified
Touchstone file, assigning a label for the cal data. A restriction on the port assignment may be defined in
addition.
'<conn_name>'
String parameters containing the name of the connector type.
'<ckit_name>'
String parameters containing the name of a calibration kit available on the
analyzer.