R&S ZVL
Measurement Examples
Basic Measurement Tasks
Operating Manual 1303.6580.32-05
106
To load a setup stored on an external device...
Connect the storage device to an USB connector.
Click Nwa-File — Recall Nwa to open the Open dialog.
Open the
Look in
drop-down list and select the storage device.
In the center of the dialog, select the folder containing the setup files.
Select a file and click
Open
to load the setup as the active setup.
Setting up a Sweep
Configurations of the different sweep types require different procedures. The following examples list the
essential settings for various sweep types.
To Set up a Lin Frequency Sweep...
Press the
Sweep
key or click
Channel - Sweep
to access the Sweep submenu.
Click Sweep Type – Lin Frequency.
Click
Channel – Center
or
Span
to access the submenu defining the sweep range.
To define a sweep range confined by a start and a stop frequency, use
Start
and
Stop.
Alternatively,
to define a sweep range of definite width around a known center frequency (e.g. an expected peak),
use
Center
and
Span.
To vary the number of measurement points (e.g. to improve the resolution of the sweep or reduce the
sweep time), click
Channel – Sweep – Number of Points.
If the active diagram is a Cartesian diagram, the x-axis is linearly scaled and labeled with the
Start
and
Stop
frequencies.
To Set up a Log Frequency Sweep...
Press the
Sweep
key or click
Channel - Sweep
to access the
Sweep
submenu.
Click Sweep Type - Log Frequency.
Click
Channel - Center
or
Span
to access the submenu defining the sweep range.
To define a sweep range confined by a start and a stop frequency, use
Start
and
Stop.
Alternatively,
to define a sweep range of definite width around a known center frequency (e.g. an expected peak),
use
Center
and
Span.
To vary the number of measurement points (e.g. to improve the resolution of the sweep or reduce the
sweep time), click
Channel - Sweep - Number of Points.
If the active diagram is a Cartesian diagram, the x-axis is logarithmically scaled and labeled with the
Start
and
Stop
frequencies.