R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
481
*RST value
ON (after
Use Default Directories
)
SCPI, Command Types
Device-specific, command or query.
Example:
See
MMEMory:STORe:TRACe:USER:CINFormation
OUTPut<Pt>...
This subsystem controls the characteristics of the analyzer’s output ports.
OUTPut<Chn>:DPORt PORT1 | PORT2
Selects a source port for the stimulus signal (drive port). The setting acts on the active trace. The effect of
the drive port selection depends on the measurement parameter associated with the active trace:
•
If an S-parameter S
<out><in>
is measured, the second port number index <in> (input port of the DUT
= drive port of the analyzer) is set equal to the selected drive port: Drive port selection affects the
measured quantity.
•
If a wave quantity or a ratio is measured, the drive port is independent from the measured
quantity:
Note:
This command is equivalent to
[SENSe<Chn>:]SWEep:SRCPort
.
<Chn>
Channel number used to identify the active trace. If unspecified the numeric
suffix is set to 1.
PORT1 ...
Test port number of the analyzer, 1 to 2.
*RST value
PORT1
SCPI, Command
Types
Device-specific, with query.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'A1'
Create channel 4 and a trace named
Ch4Tr1
to measure the wave quantity a1.
The trace automatically becomes the active trace.
OUTP4:DPOR PORT2
Select drive port 2 for the active trace.
OUTPut<Ch>[:STATe] <Boolean>
Turns the internal source power at all ports and the power of all external generators on or off.
<Ch>
Channel number. This suffix is ignored; the setting is valid for all channels.
<Boolean>
ON | OFF - Switches the power on or off.
*RST value
ON
SCPI, Command Types
Device-specific, command or query
Example:
OUTP OFF