R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
358
CALCulate<Chn>:FILTer[:GATE]:TIME:STATe <Boolean>
Determines whether the time gate for trace no. <Chn> is enabled.
<Chn>
Channel number used to identify the active trace
<Boolean>
ON – Time gate enabled
OFF – Time gate disabled
*RST value
OFF [–]
SCPI,
Command
Types
Confirmed, command or query
Example:
*RST; CALC:TRAN:TIME:STAT?; CALC:FILT:TIME:STAT?
Reset the instrument, activating a frequency sweep, and query whether the default
trace is displayed in the time domain and whether the time gate is enabled. The
response is
0;0
.
CALCulate<Chn>:FILTer[:GATE]:TIME:STOP <numeric_value>
Defines the stop time of the time gate.
<Chn>
Channel number used to identify the active trace
<numeric_value>
Stop time of the time gate.
Range [def. unit]
–99.999999999998 s to +100 s. The increment (parameters UP or DOWN)
is 0.1 ns. [s]
*RST value
+4E-009 s
SCPI, Command
Types
Confirmed, command or query
Example:
See CALCulate<Chn>:FILTer[:GATE]:TIME:STARt.
Note:
If the stop frequency entered is smaller than the current start frequency
(
CALCulate<Chn>:FILTer[:GATE]:TIME:STARt
), the start frequency is set to the stop
frequency minus the minimum frequency span
(
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN
).
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE] BPASs | NOTCh
Selects the time gate filter type, defining what occurs to the data in the specific time region.
<Chn>
Channel number used to identify the active trace
Parameters
BPASs – Band pass filter: pass all information in specified time region and
reject all else
NOTCh – Notch filter: reject all information in specified time region and pass
all else.
*RST value
BPASs
SCPI, Command
Confirmed, command or query