R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
472
SCPI, Command
Types
Device-specific, no query.
Example:
MMEM:RDIR 'C:\Documents and
Settings\NetworkService\Application Data'
Remove the specified directory.
MMEMory:SELect[:ITEM]:ALL
Includes all data subsets in the list of device settings to be stored/loaded.
*RST value
–
SCPI, Command Types
Device-specific, no query.
Example:
MMEM:SEL:ALL
Select all data subsets.
MMEMory:SELect[:ITEM]:DEFault
Sets the default list of device settings to be stored/loaded. For details on hardware settings refer to the
MMEMory:SELect[:ITEM]:HWSettings command.
*RST value
–
SCPI, Command Types
Device-specific, no query.
Example:
MMEM:SEL:DEFault
Set the default list of device settings.
MMEMory:SELect[:ITEM]:HWSettings <Boolean>
Includes the hardware settings in the list of data subsets of a device setting to be stored/loaded.
The hardware settings include:
current configuration of general device parameters (general setup)
current setting of the measurement hardware including markers
activated limit lines
A data set may include 8 limit lines at maximum. This number includes the activated limit lines
and, if available, the deactivated limit lines last used.
Therefore the combination of the non–activated restored limit lines depends on the sequence of
use with the MMEMory:LOAD:STATe command.
user–defined color setting
configuration for the hardcopy output
<Boolean>
ON | OFF – Includes the hardware settings in the list of data subsets or