R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
499
CORR:COLL:SAVE; DEL
Calculate the system error correction data and apply them to the active channel,
then delete the data.
[SENSe<Ch>:]CORRection:COLLect:METHod REFL1 | REFL2 | REFL12 | FTRans |
RTRans | FRTRans | FOPort1 | FOPort2 | FOPort12 | FOPTport | ROPTport | TPORt
| TOSM
Selects a one-port or two-port calibration type at ports 1/2.
Use the generalized command
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
to select
the calibration type for arbitrary analyzer ports or a multiport calibration type.
<Ch>
Channel number of the calibrated channel.
Parameters
Calibration types:
Normalization
(REFL1, REFL2 and REFL12 for one-port
normalization, FTRans, RTRans and FRTRans for forward, reverse, and bidirectional
two-port/transmission normalization)
, Full One Port, One Path Two Port
(FOPTport
and ROPTport are equivalent, the query returns OPT12),
TOSM. T
The numbers in the parameters denote the analyzer ports. Parameters for two-port
calibration types contain no numbers because the command is only valid for ports 1
and 2.
*RST value
–
SCPI,
Command
Types
Confirmed (with device-specific calibration types), command or query. If several
calibration types are assigned to channel <Ch>, the query returns a list of all
calibration types.
Example:
See
[SENSe<Ch>:]
CORRection:COLLect:SAVE.
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine '<cal_name>', REFL |
RSHort | FOPort | FTRans | RTRans | FRTRans | OPTPort | TOSM,
<port_no>[,<port_no>]
Selects a one-port or two-port calibration type at arbitrary analyzer ports.
<Ch>
Channel number of the calibrated channel.
'<cal_name>'
Name of the calibration (string parameter). The name serves as a reference to
delete a particular set of system correction data
(
[SENSe<Ch>:]CORRection:COLLect:DELete
).
Parameters
Calibration types:
One-port Normalization (reflection), using an open
(REFL)
or a
short (RSHort) standard, Full One Port, Two-port Normalization (
FTRans, RTRans
and FRTRans for forward, reverse, and bidirectional two-port/transmission
normalization)
, One Path Two Port, TOSM
, .
<port_no>
Port numbers of the analyzer. For an n-port calibration type (n = 1 to 2), n port
numbers must be specified. If more than n numbers are defined, the spare numbers
(the last ones in the list) are ignored. Entering less than n numbers causes an error
message.
*RST value
–