R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
415
Example:
See CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold
CALCulate<Chn>:TRANsform:DTFault:PEAK:DATA<nr>?
Returns the distance and peak response values of the detected fault no. <nr>. If the number of detected
faults (
CALCulate<Chn>:TRANsform:DTFault:PEAK:COUNt?
) is smaller than <nr>, an error
message is returned.
<Chn>
Channel number used to identify the active trace
<nr>
Current number of the fault. Faults are numbered according to their distance
(fault no. 1 has the smallest distance)
Response
Distance and peak response value of the fault, separated by a comma; see
example.
*RST value
N/a
SCPI, Command
Types
Device-specific, query only
Example:
See CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold
CALCulate<Chn>:TRANsform:DTFault:PEAK:STATe <Boolean>
Enables or disables the fault limit check.
<Chn>
Channel number used to identify the active trace.
<Boolean>
ON | OFF - Limit check on or off.
*RST value
OFF
SCPI, Command Types
Device-specific, command or query.
Example:
See CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold
CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold
Starts the automatic calculation of the number of points.
<Chn>
Channel number used to identify the active trace
Parameters
N/a
*RST value
–
SCPI,
Command
Types
Device-specific, query only
Example:
*RST; :CALC:TRAN:DTF:STAT ON
Enable distance-to-fault measurement with default settings.
CALC:TRAN:DTF:PEAK:THR –20
Specify a fault limit of –20 dB relative tor the absolute maximum of the active trace
in channel 1.