R&S ZVL
Programming Examples
Basic Tasks
Operating Manual 1303.6580.32-05
561
CALCulate1:DATA? FDATa /
Read the formatted trace data
//
Use
CALCulate<Chn>:DATA:NSWeep
to retrieve a particular trace within a group of
sweeps.
Handling of Channels, Traces and Diagram Areas
The following examples show you how to perform basic tasks related to channel and trace definition and
to the display of traces in diagram areas.
All example programs in this section have been developed and tested by means of the GPIB
Explorer provided with the network analyzer. No extra programming environment is needed.
Several Traces with Equal Channel Settings
Programming task:
Create up to four different traces with equal channel settings, assign the four 2-port
standard S-parameters to the traces and display them in up to four diagram areas.
Important remote control features for this program example
The following command sequence illustrates the structure of the remote commands discussed in section
Basic Remote Control Concepts.
In particular it shows that:
A trace can be created and handled without being displayed.
Traces are referenced by trace names. The active trace of a channel is often referenced by the