R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
546
SINGle
The settings of the active setup are reset; the name of the active setup and the
parameters of all other setups remain unchanged.
*RST value
– (*RST does not affect the preset scope setting)
SCPI, Command
Types
Device-specific, command or query
Example:
See SYSTem:PRESet
SYSTem:PRESet:USER[:STATe] <Boolean>
Selects a factory preset or a user-defined preset.
The user-defined preset can be initiated using System – Preset (manual control) or
MMEMory:LOAD:STATe.
*RST
and
SYSTem:PRESet
always initiate a factory preset.
<Boolean>
OFF - User-defined preset switched off (factory preset is used)
ON - User-defined preset switched on
*RST value
– (*RST does not affect the preset state setting)
SCPI, Command
Types
Device-specific, command or query
Example:
SYST:PRES:USER ON
Enable a user-defined preset.
SYST:PRES:USER:NAME 'C:\R_S\Instr\user\Nwa\RecallSets\Setup_2.nwa'
Select a setup file for the user-defined preset.
Press
System – Preset
(manual control)
or
use
MMEM:LOAD:STAT 1,
'C:\R_S\Instr\user\Nwa\RecallSets\Setup_2.nwa'
Carry out the use-defined preset.
SYSTem:PRESet:USER:NAME '<Setup_file>'
Specifies the name of a setup file (.nwa) to be used for a user-defined preset.
'<Setup_file>'
String parameter to specify the name and directory of the setup file to be loaded.
The default extension (manual control) for setup files is *.nwa, although other
extensions are allowed. If no path is specified the analyzer searches the default
directory (subdirectory
...RecallSets
)
*RST value
– (*RST does not affect the preset settings)
SCPI,
Command
Types
Device-specific, command or query
Example:
See SYSTem:PRESet:USER[:STATe]
SYSTem:SETTings:UPDate ONCE
Initiates an immediate update of the channel or trace settings.