R&S ZVL
Interfaces and Connectors
Rear Panel Connectors
Operating Manual 1303.6580.32-05
590
Control character
Interface function
SH1
Handshake source function (source handshake), full capability
AH1
Handshake sink function (acceptor handshake), full capability
L4
Listener function, full capability, de-addressed by MTA.
T6
Talker function, full capability, ability to respond to serial poll, deaddressed by MLA
SR1
Service request function (Service Request), full capability
PP1
Parallel poll function, full capability
RL1
Remote/Local switch over function, full capability
DC1
Reset function (Device Clear), full capability
DT1
Trigger function (Device Trigger), full capability
Interface Messages
Interface messages are transmitted to the instrument on the data lines, with the attention line being active
(LOW). They serve to communicate between controller and instrument.
Universal Commands
Universal commands are encoded in the range 10 through 1F hex. They are effective for all instruments
connected to the bus without pervious addressing.
Command QuickBASIC
command
Effect on the instrument
DCL
(Device Clear)
IBCMD (controller%,
CHR$(20))
Aborts the processing of the commands just received and sets the command
processing software to a defined initial state. Does not change the instrument
settings.
IFC
(Interface Clear)
IBSIC (controller%)
Resets the interfaces to the default setting.
LLO
(Local Lockout)
IBCMD (controller%,
CHR$(17))
The LOC/IEC ADDR key is disabled.
SPE
(Serial Poll
Enable)
IBCMD (controller%,
CHR$(24))
Ready for serial poll.
SPD
(Serial Poll
Disable)
IBCMD (controller%,
CHR$(25))
End of serial poll.
PPU
(Parallel Poll
Unconfigure)
IBCMD (controller%,
CHR$(21))
End of the parallel-poll state.
Addressed Commands
Addressed commands are encoded in the range 00 through 0F hex. They are only effective for
instruments addressed as listeners.
Command QuickBASIC
command
Effect on the instrument
GET
(Group Execute
Trigger)
IBTRG (device%)
Triggers a previously active device function (e.g. a sweep). The effect of the
command is the same as with that of a pulse at the external trigger signal input.
GTL
(Go to Local)
IBLOC (device%)
Transition to the "Local" state (manual control).